Tel:06021-58830-11

        Contact us

        06021-58830-11
        Mod:06021-58830-11
        Fax:06021-58830-0
        Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        Technology

        Nav:Home > Technology >

        What causes the tailing of HPLC column?

        High performance liquid chromatography column is composed of high pressure liquid pump separation system, detector and liquid chromatography column (including column temperature control system). The performance of liquid chromatography column is directly related to the accuracy and applicability of separation and analysis.
        What causes the tailing of HPLC column?
        1. Physical damage of column
        The physical damage of chromatographic column is the root cause of peak tailing. The solution is to replace the new column.
        2. Packing contamination in the column
        Impurities in mobile phase and sample are the main pollution sources of chromatographic column.
        At least pure solvents should be used for analysis. The water used in the mobile phase should be ultra pure water or double distilled water from all glassware. Before use, filter with 0.45um solvent microporous filter (filter) to remove the possible particles. The mobile phase is recommended to be used and prepared at present. For the solution containing salt, special attention should be paid to bacteria or precipitation. In addition, it is necessary to ensure that the container for storing mobile phase is clean.
        For complex samples, 0.45um solvent microporous filter (filter) or sample pretreatment column can be selected to pretreat the sample to ensure that there are no particulate impurities in the sample. If the sample is inconvenient to handle, a protective column should be used. When the packing in the column is polluted, the screw at the head of the column can be removed, and the contaminated filler in the front section of the column can be dug out with special tools, and then filled with the same filler again to repair. Or wash the chromatographic column (about 20 to 30 times the volume of the column, or as the case may be; in addition, the detector can not be connected at this time), wash the pollutant out of the column, and then use it according to the direction indicated by the column.
        3. There are foreign matters at the column inlet
        When it is determined that there is foreign matter at the column inlet, remove the screw on the column head, take out the filter cap and put it in 20% nitric acid solution, clean it with ultrasonic wave for about 20 minutes, then put it in ultra pure water, and clean it with ultrasonic wave for about 10 minutes, and then put it into the chromatographic column again.
        4. The sample concentration is too high to cause column overload
        The overload of the sample on the column can cause the peak broadening and tailing (or tongue stretching).
        The adverse effect can be eliminated by appropriately reducing the injection volume or sample concentration (the sensitivity of the detector can be increased if necessary) until the peak shape and retention time are no longer changed. The resolution can also be improved by reducing the injection volume. Under normal conditions, the injection volume of each compound in the sample is maintained in the range of 3-50ug in a 150 × 4.6mm column, which will not cause obvious overload.
        5. The sample solvent is not correct
        Choose the appropriate sample solvent to eliminate unnecessary interference. The mobile phase should be used to dissolve the sample.
        6. Out of column effect
        The out of column effect (i.e., the pipeline between injection valve, chromatographic column and detector is too long, the diameter is too large, the pipe joint does not match, and there is dead volume) is one of the main factors affecting the column efficiency. Therefore, under possible conditions, the connecting pipeline at both ends of the column should be as short as possible, the inner diameter of the connecting tube should be as small as possible, the incision must be smooth and smooth, and the dead volume should be reduced as much as possible, so as to prevent the situation that the real column efficiency can not be reflected due to the sample diffusion.
        7. Insufficient or inappropriate buffer
        In the separation with poor buffer or low ionic strength, the retention time reproducibility is poor and the peak shape is tailed. Increasing the buffer concentration (matching the sample size) can improve this situation.
        8. Silanol group action
        A careful analysis of the surface properties of the packing in the column shows that the surface of the packing material is about half covered by the bonding phase, and the rest is the unconjugated silanol group. The so-called retention is the result of the interaction between the sample molecules and the bonding phase. However, the interaction between acidic or basic compounds and the residual silanol groups or metal impurities on the surface of silica gel leads to double retention mechanism, thus leading to peak tailing.
        In order to reduce the peak tailing, 25mm triethylamine (inhibitor) can be added into the mobile phase. Triethylamine has strong interaction with silanol group, which can reduce or block the interaction between sample molecule and silanol group, so as to ensure the normal retention mechanism and greatly alleviate the peak shape tailing. Using long chain silyl alcohol group inhibitors, although the effect is slow, but the persistence is longer than triethylamine. In addition to the polar interaction between the amino group and the silanol group, the nonpolar part of the macromolecule and the stationary phase will also react in reverse phase, resulting in the retention phenomenon. If the inhibitor is added to the sample, the effect will be significant.
        9. Metal contamination in the column
        Metal contamination (Fe, Ni, etc.) in the column can cause the tailing of some compounds. The metal can be brought in by the packing itself, or the metal filter at the inlet of the column can be slowly dissolved by the corrosive mobile phase and deposited in the column packing with the mobile phase. For example, after the C18 column packing is contaminated by metal, the acid / alkaline sample tailing can be caused. The packing can be bonded after alkaline washing / pickling, and the peak obtained is sharp and symmetrical.
        For more related technologies, please consult MS technologies technicians
        Contact Feedback
        Tel:06021-58830-11
        Sitemap( Google
        Copyright:MS Technologies Tel:06021-58830-0  Fax:06021-58830-11    Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        MS Technologies

        主站蜘蛛池模板: 精品国产乱码久久久久久浪潮 | 性按摩xxxx| 亚洲人成色777777在线观看| _妓院_一钑片_免看黄大片| 日本高清在线中文字幕网| 亚洲欧美色一区二区三区| 老子影院午夜伦不卡| 天天操天天干天天做| 久久国产免费一区二区三区| 精品亚洲一区二区三区在线播放| 国产成年无码v片在线| 99久久精品九九亚洲精品| 探花视频在线看视频| 么公的又大又深又硬又爽视频| 狠狠操.com| 哒哒哒免费视频观看在线www| 国产色丁香久久综合| 国色天香精品一卡2卡3卡| 中国特级黄一级**毛片| 日韩免费a级在线观看| 亚洲国产精品无码久久青草| 男人j桶进女人p无遮挡在线观看| 国产v片成人影院在线观看| 国产自产21区| 国产精品爆乳奶水无码视频| blacked欧美一区二区| 扒开双腿猛进湿润18p| 久久精品成人一区二区三区| 精品xxxxxbbbb欧美中文| 国产精品久久久福利| a级毛片免费观看网站| 成人妇女免费播放久久久| 久久精品国产96精品亚洲| 欧美巨大xxxx做受中文字幕| 伊人a.v在线| 精品影片在线观看的网站| 国产中文在线观看| 黑人极品videos精品欧美裸| 国产精品永久免费视频| 99国产精品国产精品九九| 好吊色永久免费视频大全|