        How to operate the cleaning and regeneration of HPLC column?

        In general, for the commonly used reversed-phase chromatographic column, if the column efficiency is seriously reduced, it is necessary to consider cleaning and regeneration. It is recommended that the following process be used for washing:
        [flushing process]
        20 times column volume water: acetonitrile = 95%: 5% (used to remove buffer salt in the system, and the proportion can be adjusted according to the concentration of common buffer salt)
        Acetonitrile 20 times column volume
        Isopropanol 5 times column volume
        N-hexane with 20 times column volume
        Isopropanol 5 times column volume
        Acetonitrile 20 times column volume
        20 times column volume water: acetonitrile = 95%: 5% (the same proportion can be maintained as step 1)
        Rebalance the column with the required mobile phase ratio
        As for the regeneration of chromatographic column, I believe you can see a lot of related articles. It is suggested that you can choose the regeneration conditions according to the actual needs.
        What Elsa is going to tell you today is about the volume of the column in the conditions given above.
        [split arrow dynamic]
        In fact, when the chromatographic column is washed, many laboratories are used to use the washing time to evaluate whether the desired effect is achieved. For example, a master told you to flush the pillar for half an hour.
        However, the volume of washing solvent should be used to evaluate the washing effect. For example, 5 times the column volume of isopropanol, 20 times of column volume of acetonitrile, etc., what do these column volumes refer to?
        In fact, the volume of the column needs to be calculated according to the inner diameter and length of the column you use.
        It is easy to calculate the column volume by the formula. For example, the volume of a column with 4.6 mm inner diameter and 150 mm column length is as follows:
        If you want to wash the column with 20 times the volume of acetonitrile, you need about 34 ml of acetonitrile. If your column flow rate is 1ml / min, you need to rinse with acetonitrile for 34 minutes.
        When the column flow rate was 1.5 ml / min, it was necessary to rinse with acetonitrile for about 23 minutes.
        I don't know. Can you understand that?
        Finally, we should remind you that when washing the chromatographic column, do not connect the detector at the end of the column, but a container that can store the waste liquid.
        For more related technologies, please consult MS technologies technicians
        Contact Feedback
        Tel:06021-58830-11
        Sitemap( Google
        Copyright:MS Technologies Tel:06021-58830-0  Fax:06021-58830-11    Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        MS Technologies

        主站蜘蛛池模板: 久久久久香蕉视频| 免费在线成人网| 97精品国产91久久久久久| 日本一区二区三区免费观看| 亚洲成a人片在线观看中文| 精品欧美一区二区三区久久久| 国产福利拍拍拍| aⅴ一区二区三区无卡无码| 日本三级免费观看| 亚洲人成网站在线观看播放| 99久久精品这里只有精品| 日本高清黄色片| 亚洲毛片av日韩av无码| 美国成人a免费毛片| 国产成人精品免费久久久久| 99久久国产热无码精品免费| 扒下胸罩揉她的乳尖调教| 亚洲av无码专区在线播放| 热带雨电影无删减在线播放| 四虎成人精品在永久在线观看| jizz国产丝袜18老师美女| 日本肉体xxxx裸交| 亚洲国产美女精品久久久久| 精品一区二区三区波多野结衣 | 男人j桶女人j免费视频| 国产农村乱子伦精品视频| 尤物视频在线看| 天天做天天爱天天综合网2021| 中文字幕日韩欧美一区二区三区| 最新无码a∨在线观看| 亚洲欧美日韩色| 稚嫩娇小哭叫粗大撑破h| 国产主播在线一区| 免费黄网站大全| 在线观看亚洲一区| 一区二区三区精品视频| 无遮挡韩国成人羞羞漫画网站| 九九久久精品国产AV片国产| 欧美日韩中文视频| 亚洲综合精品伊人久久| 精品亚洲麻豆1区2区3区|