        The most complete maintenance method of liquid chromatograph

        Common problems in the use of liquid chromatography column
        Liquid chromatography (LC) is an efficient analytical method, and the column is the "heart" of chromatographic analysis system. Due to the variety of samples to be analyzed and the complex sample matrix, the chromatographic column is easily contaminated by strong retention substances. These strong retention substances are usually difficult to elute. After several times of sample loading, they gather at the end of the column, block the sieve plate, and even interact with the filler, resulting in the denaturation of the packing, which will lead to the decrease of column efficiency, the decrease of resolution and the abnormal peak pattern. Therefore, it is necessary to maintain the column.
        Maintenance of Yuexu technology liquid chromatography column
        The chromatographic column needs to be maintained, but different manufacturers have different maintenance methods. Yuexu technology liquid chromatography column is recommended to be used in forward direction and back flush. Because the aperture of the sieve plate used at both ends of Yuexu's chromatographic column is the same, there will be no back flushing out the packing. When the sample analysis is finished, most of the pollutants are concentrated in the end of the column. At this time, if the backwash is carried out immediately, the service life of the column packing can be greatly prolonged.
        How to recoil liquid chromatography column
        The principle of column backwash in liquid chromatography is to wash with solvent with stronger elution capacity than mobile phase.
        The reversed-phase chromatographic column (C18, C8) can be reversed washed with less polar solvent
        Water → methanol → acetonitrile → isopropanol → dichloromethane → isopropanol → acetonitrile
        The positive phase chromatographic column (SiO2, NH2, CN, diol) can be backwashed with solvent with larger polarity
        Hexane → dichloromethane → ethyl acetate → isopropanol → acetonitrile → methanol
        If there is buffer salt in the mobile phase and the column pressure rises abnormally during the operation sequence, the buffer salt will precipitate on the column bed. The following maintenance methods are recommended:
        Wash 60 column volumes with methanol / water = 10 / 90 at 35 ℃ and column temperature;
        Wash the column with methanol / water = 10 / 90 at a flow rate of 0.2ml/min overnight.
        For more related technologies, please consult MS technologies technicians
        Contact Feedback
        Tel:06021-58830-11
        Sitemap( Google
        Copyright:MS Technologies Tel:06021-58830-0  Fax:06021-58830-11    Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        MS Technologies

        主站蜘蛛池模板: 国产精品videossex另类| 男女午夜爽爽大片免费| 怡红院美国分院一区二区| 亚洲小说区图片区另类春色| 韩日午夜在线资源一区二区| 性刺激久久久久久久久| 亚洲av日韩综合一区久热| 老师您的兔子好软水好多动漫视频| 女人被男人躁的女爽免费视频| 亚洲小说区图片区另类春色| 精品国产理论在线观看不卡| 国产精品自产拍在线观看花钱看| 久久婷婷激情综合色综合俺也去| 精品久久久中文字幕一区| 国产精品午夜电影| 一个人看的日本www| 欧美大片在线观看完整版| 国产中文字幕视频| 120秒男女动态视频免费| 无遮挡无删动漫肉在线观看| 亚洲视频在线免费播放| 色吊丝二区三区中文字幕| 在线精品自拍亚洲第一区| 九九在线中文字幕无码| 精品人妻潮喷久久久又裸又黄| 国产精品社区在线观看| 一区二区三区四区视频在线 | 在线天堂中文www官网| 亚洲H在线播放在线观看H| 美女航空一级毛片在线播放| 国产韩国精品一区二区三区久久| 久久精品一区二区三区中文字幕| 欧美激情视频一区二区| 免费a级毛片18以上观看精品 | 国产超爽人人爽人人做| ww美色吧com| 最新在线中文字幕| 免费大片av手机看片| 国产成人精品啪免费视频| 国产综合欧美日韩视频一区| 中文字幕一区二区人妻性色|