Tel:06021-58830-11

        Contact us

        06021-58830-11
        Mod:06021-58830-11
        Fax:06021-58830-0
        Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        Technology

        Nav:Home > Technology >

        Use and maintenance of chromatographic column

        Maintenance of chromatographic column
        1. A pre column was used to protect the analytical column (silica gel has a certain solubility in polar / ionic mobile phase).
        2. The mobile phase must be degassed and filtered before use.
        3. Avoid drastic changes in composition and polarity of mobile phase.
        4. The pH stability range of most RP-HPLC columns is 2-7.5, which should not exceed the pH range of the column.
        5. If polar or ionic buffer solution is used as mobile phase, the column should be washed clean after the experiment and stored in acetonitrile.
        6. Pressure rise is a signal that the precolumn needs to be replaced or the chromatographic column washed.
        System considerations
        1. When opening and closing the system, the flow rate should be buffered (1ml / min → 0.8ml/min → 0.4ml/min → 0ml / min), so as to avoid the collapse of column filling due to sudden pressure change.
        2. The mobile phase is the buffer system. Do not switch to strong solvent directly. Sudden conversion to high concentration organic solvent may precipitate the buffer solution in the HPLC flow system, which will lead to more problems, such as column head blockage, pipeline blockage, pump leakage, piston damage or injection valve spindle failure. The strong solvent should be replaced after washing 5-10 column volumes with buffer free mobile phase (i.e. replacing buffer with water).
        3. Do not wash the column directly with 100% organic solvent after the test. Wash the column with 90% water-10% organic solvent (if the mobile phase is buffer solution, pure water or 5% organic solvent water solution) for at least 30 min, then wash the column with organic solvent for about 30 min, and finally store the column in organic solvent.
        4. Before replacing the chromatographic column, the chromatographic column must be washed and balanced, and then stored in organic solvent (preferably in acetonitrile).
        5. When removing the chromatographic column, both ends should be sealed immediately.
        6. Except for special cases, the pre protection column should be used in general.
        The reason of equilibrium chromatographic column
        The reversed-phase chromatographic column is stored in acetonitrile / water after delivery test. As the chromatographic column may dry out during storage or transportation, if it is not balanced, some salts, lipids, fatty substances, humic acid, hydrophobic protein and other biological substances easily interact with the HPLC column, so that these substances can be adsorbed on the chromatographic column, Contaminate the column.
        How to balance the column?
        Therefore, before using mobile phase to analyze samples, we should first use methanol or acetonitrile equilibrium chromatographic column with 10-20 times of column volume; if the mobile phase we use contains buffer salt, we should pay attention to "transition" with pure water. During the equilibration process, the flow rate is increased slowly and the column is equilibrated with mobile phase until a stable baseline is obtained (if the buffer salt or ion pair reagent level is low, it will take a long time to equilibrate).
        What might happen after column contamination?
        The sample components retained in the chromatographic column are accumulated to a certain extent to make them start to form a new stationary phase. Sometimes these analytes can interact with these impurities to form a certain separation mechanism. The interference of these non target products can be detected by the detector, and can form chromatographic peaks, bubbles, baseline disturbance and baseline drift, and "strange peaks" appear on the chromatogram ”The results showed that negative peak, wide peak, steamed bread peak, double headed peak, trailing peak and retention time fluctuated. These behaviors are usually found only through parallel experiments.
        For more related technologies, please consult MS technologies technicians
        Contact Feedback
        Tel:06021-58830-11
        Sitemap( Google
        Copyright:MS Technologies Tel:06021-58830-0  Fax:06021-58830-11    Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        MS Technologies

        主站蜘蛛池模板: 亚洲av无码一区二区三区鸳鸯影院| 国产一区二区三区不卡av| 丁香花免费高清视频完整版 | 欧美精品高清在线观看| 国产一二三区视频| 1024视频在线| 天堂а√在线中文在线| 中文字幕日韩精品无码内射| 校花主动掀开内裤给我玩| 亚洲视频免费在线看| 色多多在线观看| 国产福利一区二区三区在线视频| hd日本扒衣党视频播放| 无遮挡呻吟娇喘视频免费播放| 亚洲人成在线播放网站| 特级aaaaaaaaa毛片免费视频| 国产69精品久久久久9999 | 今天免费中文字幕视频| 色噜噜狠狠狠综合曰曰曰| 国产福利免费看| 94久久国产乱子伦精品免费| 思思99re66在线精品免费观看| 久久天天躁狠狠躁夜夜免费观看| 欧美日韩亚洲一区| 免费a级毛片无码| 羞羞视频免费观看| 国产成人手机高清在线观看网站 | 蜜桃精品免费久久久久影院| 国产精品一区二区久久不卡| 99久久国产免费中文无字幕| 成人亚洲欧美日韩中文字幕| 久久人人爽人人爽av片| 欧美40老熟妇| 亚洲欧美另类综合日韩| 男人女人真曰批视频大全免费观看| 国产99久9在线视频| 香蕉污视频在线观看| 国产精品久久久久影院| 99久久精品九九亚洲精品| 婷婷色香五月综合激激情| 中文字幕欧美日韩高清|