        Column installation, commissioning and maintenance

        1. Matching of column head type and stainless steel capillary joint
        Chromatographic column is a consumable, not the original equipment of the instrument. If the depth of the joint and the column head do not match, there will be leakage or excessive dead volume. The depth of the joint is longer than that of the column head, so it is not easy to tighten and leak liquid; the depth of the joint is shorter than that of the column head, and there is a gap in the column head, resulting in the birth and death volume, broadening the spectrum band and tailing the peak.
        2. Matching conversion of solvents
        If the solvent does not match with the mobile phase, it should be converted before use. Especially when the mobile phase contains buffer salt, if the preservation solvent is pure organic phase or organic phase is higher, if the new column is used directly, the buffer salt will crystallize and precipitate in the column, and the new column will be permanently and irreversibly damaged. The preservation solvent of normal phase column is n-hexane. If it is to be converted to HILIC column mode, dichloromethane or ethyl acetate should be used during the conversion due to the poor mutual solubility of n-hexane and methanol acetonitrile.
        3. Balance and aging of new column before use
        Generally, the balance has been carried out in the factory inspection, but the time from the chromatographic column to the end user is different. It is better to rebalance the chromatographic column before the formal determination by the user. The best way to balance and age together is to run a complete analysis program (including injection) several times until the stability of peak shape, retention time and peak area is observed. The so-called "aging" is a term borrowed from the gas phase in order to achieve the adsorption saturation of analytes in the chromatographic column and the whole flow path of the liquid phase system. For some specific analytes, such as those with molecular weight greater than 1000, the aging time will be longer due to the slow diffusion speed, so the aging can be accelerated by large concentration injection or continuous injection of multiple needles in the same elution cycle.
        4. PH application range
        It is generally considered that the pH range of silica gel matrix column is 2-8, which is very rough. The type of silica gel, the temperature of use, the type of stationary phase bonded on the surface of silica gel, and the buffer salt have influence on this. The pH tolerance range of silica gel is larger than that of fillers with small pore volume and high bonding density. Some use high-quality silica gel, high-density bonding and double sealing, or special process can increase the pH tolerance range to 10. Phosphate buffer salt has strong permeability, and has the side effect of accelerating the dissolution of silica gel. Organic hybrid silica gel and silica gel filler coated with hybrid layer on the surface of silica gel can be used in the pH range of more than 12 due to the existence of organic matter. Read the instruction manual at the first time after getting the column to confirm the sealing solution in the column. If possible, column efficiency test can be carried out according to COA report.
        5. Column preservation
        For short-term storage (overnight or every weekend), it is better to use the mobile phase (without buffer salt) to minimize the balance time of the next use. It is generally recommended to use pure methanol or acetonitrile for long-term storage of reversed-phase columns. On the one hand, pure organic phase preservation can minimize the hydrolysis of bonding phase, but pure methanol (acetonitrile) will elute the bonded phase which has been hydrolyzed and temporarily adsorbed in the column, thus speeding up the loss process of the stationary phase and shortening the service life of the stationary phase. The pure organic phase has the disadvantage of easy volatilization and drying of the column bed. The use of about 80% organic phase is also a good choice, and can avoid the growth of bacteria. It is strongly recommended to label the column with the preservation solvent.
        The CN based column is unstable in organic polar solvents (which may lead to the collapse of the column bed structure) and is suitable for low temperature storage in aqueous phase. The column bed should not be damaged due to freezing of stationary phase. 0.05% sodium azide solution can be added to the chromatographic column suitable for preservation in aqueous solution, such as ion exchange column and water-based SEC column to prevent bacterial growth.
        It is recommended that the positive phase column be preserved in the mobile phase used, whether in the short term or in the long term.
        For more related technologies, please consult MS technologies technicians
        Contact Feedback
        Tel:06021-58830-11
        Sitemap( Google
        Copyright:MS Technologies Tel:06021-58830-0  Fax:06021-58830-11    Add:Zeppelinstr. 5, 63741 Aschaffenburg, Germany

        MS Technologies

        主站蜘蛛池模板: 伊人热人久久中文字幕| 杨幂一级做a爰片性色毛片| 国产大学生粉嫩无套流白浆| gogogo高清在线观看中国| 欧美日韩国产亚洲一区二区三区| 国产91久久久久久久免费| eeuss影院www在线观看免费| 欧美人与禽交另类视频| 国产亚洲视频在线播放大全| 91chinese在线| 岳代理孕妇在线风间由美| 久久精品国产精品国产精品污| 精品午夜久久福利大片免费| 国产精品色午夜免费视频| 久久天天躁狠狠躁夜夜不卡| 精品人妻少妇一区二区| 国产成人一区二区三区免费视频 | 特黄特色一级特色大片中文| 国产乱码一区二区三区| 亚洲另类专区欧美制服| 国精品午夜福利视频不卡| 久久精品中文字幕一区| 欧美精品va在线观看| 免费观看国产网址你懂的| 3d性欧美动漫精品xxxx| 天天射天天爱天天干| 久久精品国产免费一区| 真正全免费视频a毛片| 国产精品国产三级在线专区| 中日韩精品视频在线观看| 波多野结衣在线影院| 向日葵app在线观看下载大全视频| 国产精品嫩草影院人体模特| 成人免费视频88| 亚洲国产高清美女在线观看| 粉色视频下载观看视频| 国产一区在线电影| 2020国产精品永久在线| 女人洗澡一级特黄毛片| 中文字幕免费在线观看动作大片 | 国产精品成人免费视频网站|